PXIe-5842 Unable to Complete Self-Calibration Over MXI Gen 1x1

Updated Oct 19, 2026

Reported In

Hardware

  • PXIe-5842
  • PXIe-8361
  • PXIe-8360
  • PCIe-8361
  • PCIe-8362
  • PCIe-8363
  • PXIe-8364

Issue Details

I am unable to perform self-calibration on my NI PXIe-5842. The self-calibration process fails when the device is connected through an MXI Gen 1x1 link.
During self-calibration, the following error is reported: "The transfer did not complete within the timeout period"


The issue is reproducible whenever self-calibration is attempted using the MXI Gen 1x1 connection.

Solution

This behavior will occur when using an MXI Gen 1x1 link (PXIe-836x/PCIe-836x devices) with the NI PXIe-5842 during the self-calibration process. Please note that MXI Gen 1x1 is an unsupported configuration with the PXIe 5842 VST as the VST is a PCIe Gen3x8 configuration.
 
To resolve the issue:
  1. Use a Gen 3 MXI link (PCIe 8398 and PXIe 8398) for the self-calibration procedure.
  2. Use an embedded controller instead of the MXI Gen 1x1 connection.
  3. Re-run self-calibration after changing to one of the supported controller configurations.