I purchased a PXIe-5842 together with a PXIe-5655 as part of a bundle. When I received the hardware, both calibration certificates displayed the same Calibration ID. I want to know whether the two modules are tied together to achieve their maximum performance or if they can be used independently.

Solution

The PXIe-5655 and PXIe-5842 are calibrated individually, even when they are purchased together as a bundle and share the same Calibration ID. You may use the PXIe-5655 and PXIe-5842 independently as required by your application.
A shared Calibration ID on bundled hardware does not indicate that the devices must remain together during operation. Calibration is performed on each individual module.

Additional Information

A shared Calibration ID may be used for administrative or documentation purposes when hardware is shipped as a bundle. This does not affect the independent calibration status of the individual modules.