Solution
NI-9207 uses a Delta-Sigma ADC architecture with a multiplexer (MUX). The specified 500 S/s corresponds to a conversion time of 2 ms per sample.
- Use 500 S/s as the aggregate module sampling rate.
- Divide the aggregate rate by the number of enabled channels to determine the effective per-channel sampling rate.
Examples:
- One channel enabled:
- Two channels enabled:
- 16 channels enabled:
- 31.25 S/s per channel (500 รท 16)
The effective per-channel sampling rates already account for the conversion process and channel multiplexing behavior.
The specified 500 S/s (2 ms conversion time) represents the time required to acquire a single sample through the Delta-Sigma ADC measurement chain. This timing includes:
- Oversampling
- Digital filtering
- Decimation
- Multiplexer (MUX) switching and associated delay
Therefore, the published 500 S/s specification includes the channel-to-channel delay associated with the module's conversion process.