Error -1074102151 When Self-Calibrating NI PXIe-4162 in NIMAX

Updated Jul 22, 2026

Reported In

Hardware

  • PXIe-1092
  • PCIe-8383
  • PXIe-8381
  • PXIe-4162
  • PXIe-4163

Issue Details

I am using Advantech IPC 788E, along with NI PCIe-8383 to remote control a NI PXIe-4162 and a NI PXIe-4163 SMU in the NI PXIe-1092 chassis thru the NI PXIe-8381.

The NI PXIe-4162 is slotted in slot 2 and the NI PXIe-4163 is in slot 7 in the chassis and when I try to self calibrate the NI PXIe-4163 in slot 2. it returns an error -1074102151, while the SMU in slot 7 pass the self calibration.