When I try to open an ATML that is written in TD Schema v3.0. I get Error -17901, No converters can conver this file to a sequence file. I can open ATML files written in v1.0.
Update ATML Toolkit: Download and install TestStand ATML Toolkit 2017. This version officially supports the ATML 3.0 (IEEE 1671.1:2017) schema.
Configure Translator Settings: The ATML version must be explicitly defined in the translator configuration.
Launch the TestStand Sequence Editor.
Go to Tools > Configure ATML TD Translator....
Open the Schemas tab.
Check Test Description Schema V3.0.
Click OK and try importing the TD file again.