PXIe-6593 Digital I/O Connector Output

Updated Dec 18, 2025

Reported In

Hardware

  • PXIe-6593

Issue Details

The specifications of the PXIe-6593 module state the following:

ConnectorMolex™ Nano-Pitch I/O™
5.0 V Power±5%, 50 mA maximum, nominal

What does this mean?

 

Solution

The 5 V pin on the connector is generated by an on-board charge-pump supply, regulated to ±5 %. It is designed to provide up to 50 mA total output current for light external biasing or logic reference purposes. Short inrush currents up to approximately 100 mA are tolerated, but continuous loading beyond 50 mA should be avoided.