Using 32 AI Channels of USB-6363 with BNC Termination

Updated Oct 17, 2024

Issue Details

  • What is the pin assign of USB-6363 with BNC Termination (PN: 782258-01) for Single-end Measurement?
  • How can I use 32 AI channels of USB-6363 with BNC Termination (PN: 782258-01)?

Solution

USB-6363 with BNC Termination (PN: 782258-01) has 16 BNC channels. The pin assign of AI 0 - 7 and 16 - 23 are the same with other type of USB-6363 as following.

 

6363terminal2.png

 

For example, AI0 (BNC) consists of AI 0 (AI 0+ of the BNC) and AI 8 (AI 0- of the BNC). AI16 (BNC) consists of AI 16 (AI 16+ of the BNC) and AI 24 (AI 16- of the BNC).

Using 32 AI channels for single-end measurement, you need to have BNC to wire lead converter of some other things to use BNC + and - channel separately.