Missing the "Driver" Option for USB3 Vision Cameras in NI MAX on Windows 11

Updated Sep 30, 2026

Reported In

Software

  • Measurement & Automation Explorer (MAX)

Driver

  • NI-IMAQdx

Issue Details

I have a USB3 Vision camera which I am using with the NI IMAQdx driver to acquire images. Recently I have upgraded my Microsoft Windows© version from Windows©10 to Windows©11, and I cannot change the camera driver by right clicking the camera like I used to for Windows 10. 

For example, I have a Basler USB3 Vision camera. When using it with Windows 10, by right-clicking the camera I can select the driver I want: 

image.png

However, when using Windows 11, the driver option is missing: 

image.png

How can I solve this issue?

Solution

This is a known OS-specific issue, with Bug report number 2864549, it has been resolved as of Vision Acquisition Software (VAS) 2025Q1 version. Any version before this one may still present the behavior. It is recommended upgrading your VAS software.

 

See the workaround below showing how to change drivers in the Device Manager in case upgrading the VAS software is not an alternative: 
 

  1. Open the Windows Device Manager
  2. The camera should be visible under the manufacturer's tree. Select Update Driver
    • image.png
  3. Select Browse my computer for drivers,
    • image.png
  4. Select Let me pick from a list of available drivers,
    • rtaImage (2).jpg
  5. Select NI-IMAQdx USB3 Vision Device.
    • image.png

​​​​​After performing the above steps, the driver for the selected camera will change, and you will be able to use it with NI software.