Maximum Time Width of 200 MV Output From PXI-6733 DAC When Power Is Turned On

Updated Jun 3, 2024

Reported In

Hardware

  • PXI-6733

Issue Details

As the power-on state of PXI-6733 is stated as 0V (±200 mV), what is the maximum time width of the 200mV output from the PXI-6733 DAC when the power is turned on?
Is it within 1 second at the maximum, may output more than 10 seconds, indefinite depending on conditions, etc.?

Solution

If it is regarding a random error of ±200 mV at the power-on state, there is no clear confirmation about how long the time width.
Whenever PXI-6733 is powered and the random ±200 mV appears on its output it will stay there until a value is given to it.

Additional Information

PXI-6733 is not the best fit if the application uses very sensitive DUT.
If the PXIe-6733 causes random error of ±200 mV at the power-on state, then the solution is to change the PXI-6733 with some other similar hardware or can add relays in front of the PXI-6733 and after initialization of the system switch on the relays (something similar to NI SMUs or AWGs application).