Cannot Add NI Veristand Engine on cRIO-9068

Updated Sep 28, 2023

Reported In

Software

  • VeriStand 2021 R2
  • LabVIEW 2021

Issue Details

I am using cRIO-9068 with VeriStand 2021 R2, when I try to install NI VeriStand Engine into the cRIO target, the option is not available.
1.jpg
 
 

Solution

VeriStand 2021 or later only supports 64-bit Linux RT target, cRIO-9068 is running on 32-bit ARM Linux RT so it's not supported. You also can refer to the VeriStand 2021 R2 Readme:
 
2.png

To use VeriStand with cRIO-9068, use VeriStand and LabVIEW version 2020 or the lower version.