GPIB Can Communicate in NI MAX but Fail in LabVIEW Shipping Example

Updated Sep 28, 2023

Reported In

Hardware

  • GPIB-USB-HS+

Software

  • Measurement & Automation Explorer (MAX)
  • LabVIEW

Driver

  • NI-488.2

Other

Agilent E5061B Network Analyzer

Issue Details

I tried to run LabVIEW Shipping Example Simple GPIB.vi but pop out error 7: Nonexistent GPIB board. 
 1.png
 
I tested the communication with the GPIB instrument in NI MAX by sending *IDN? command and return product information successfully.
2.png

Solution

  1. Make sure the GPIB address is the only number for this case. You may check the IP address in NI MAX.
    •   3.jpg
  2. Restart the PC and run the shipping example again.
  3. Enable Tulip passport in NI MAX. Refer to VISA Tulip Passports Are No Longer on My System - NI