在Calibration Executive上校准PXI-4461时出现错误-200112

更新 Dec 28, 2023

适用于

硬件

  • PXI-4461
  • PXI-4462

软件

  • Calibration Executive

驱动

  • NI-DAQmx

问题详述

我正在尝试使用Calibration Executive来校准PXI-4461。在预热步骤中,我收到以下错误消息:

Error -200112 occurred at DAQmx Initialize External Calibration.vi: 3390001.

Possible reason(s):
Calibration handle is invalid.
Open a calibration session to get a valid calibration handle. Use the valid calibration handle obtained when the calibration session was opened.

错误200112.png

解决方案

该错误是由CAR 1149063上报告的错误引起的。该错误将影响PXI-4461和PXI-4462。要解决该问题,请安装NI-DAQmx 20.7 ,该版本已包含错误修复。