Error -200112 When Calibrating PXI-4461 on Calibration Executive

Updated May 24, 2021

Reported In

Hardware

  • PXI-4461
  • PXI-4462

Software

  • Calibration Executive

Driver

  • NI-DAQmx

Issue Details

I'm trying to use Calibration Executive to calibrate a PXI-4461. During the warm-up step, I receive the following error: 

Error -200112 occurred at DAQmx Initialize External Calibration.vi: 3390001.

Possible reason(s):
Calibration handle is invalid.
Open a calibration session to get a valid calibration handle. Use the valid calibration handle obtained when the calibration session was opened.


error 200112.png

Solution

The error is caused by a bug reported on CAR 1149063. The error will affect both the PXI-4461 and the PXI-4462. To resolve the issue, install NI-DAQmx 20.7 as the bug fix is included in this version.