Multi-Module RF Instrument Trigger Source

Updated Jun 15, 2023

Reported In

Hardware

  • PXI-5661
  • PXIe-5663
  • PXIe-5665
  • PXIe-5667
  • PXIe-5668
  • PXI-5670
  • PXI-5671
  • PXIe-5672
  • PXIe-5673
  • PXI-5142
  • PXIe-5622
  • PXIe-5624
  • PXI-5421
  • PXI-5441
  • PXIe-5442
  • PXIe-5450
  • PXIe-5451

Driver

  • NI-RFSA
  • NI-RFSG

Issue Details

Which submodule of a multi-module instrument shares and receives the trigger signals on the PXI Backplane?

Solution

To locate this information you need to take look at the documentation of the multi-module instrument for example the PXIe-5667. If you look through the PXIe-5667 Features, you can find the description of all individual modules. Also, here you can find the block diagrams which depict if the individual module manages the trigger sharing functionality. Indicators for this are the PXI_CLK100, the PXI_Trig lines, and the front panel PFI connectors.


PXIe-5622
PXIe-5622 Block Diagram

The following table shows you which RFSA and RFSG devices use which submodule to export and receive trigger signals from the PXI Backplane:
 
TypeHardwareTrigger Source
RFSAPXI-5661PXI-5142
PXIe-5663/5663EPXIe-5622
PXIe-5665PXIe-5622
PXIe-5667PXIe-5622
PXIe-5668PXIe-5624
RFSGPXI-5670PXI-5421
PXI-5671PXI-5441
PXI-5672PXIe-5442
PXIe-5673/5673EPXIe-5450/5451
 

Additional Information