SW : MPManager 3.x
HW : MP500
When running an Analog Measurement script on MPManager, I get the Error 540 in the log pannel.
You will need to send the MP500 you are using to NI in order to perform a calibration of the device.Please contact NI technical support team on NI.com when opening a Service Request case using the NI Service Request Manager (SRM).To open a service request on NI.com please follow the steps described at this link.
This issue can happen in the following case :
- If the last calibration date is too old.
- If the Analog Measurement license has been added after the last calibration was performed.
What Advanced Analog Measurement Feature of MP500 Stands For?