Differential Measurement with PXIe-5105

Updated Feb 26, 2019

Reported In

Hardware

  • PXIe-5105
  • PXIe-5160
  • PXIe-5162
  • PXIe-5122
  • PXIe-5164
  • PXIe-5170
  • PXIe-5171
  • PXIe-5172
  • NI-5191

Issue Details

I would like to do differential measurement with my PXIe-5105 oscilloscope card, but I don't find any related hints in the manual of the product. Is there a possibility to conduct differential measurements with the PXIe-5105?
 

Solution

Yes, it is possible to measure voltage signals differentially with the PXIe-5105 oscilloscope card by applying the NI 5191 differential probe. Since the PXIe-5105 oscilloscope card has SMB connectors, and the NI 5191 active differential probe has BNC output, a BNC to SMB adapter will be necessary.