Can I Use RJ45 Connector With RM-4339?

Updated Dec 25, 2018

Reported In

Hardware

  • PXIe-4339
  • RM-4339

Issue Details

I have a PXIe-4339 and RM-4339. Can I use a connector RJ45 instead of RJ50 to connect my strain gage?

 

Solution

NI does not recommend using RJ45 connectors due to the possibility of damaging the outer pins used for shunt calibration.
NI recommends RJ50 connectors. 
More information can be found on the User Guide and Terminal Block Specifications NI PXIe-4339 and RM-4339 .
 
 

WAS THIS ARTICLE HELPFUL?