DMM Current and Resistance Measurement Linearity

Updated Mar 7, 2018

Reported In

Hardware

  • PXIe-4080
  • PXIe-4081
  • PXIe-4082

Issue Details

I reviewed the specifications of my DMM and linearity is listed only in the DC Voltage section, but I need to perform different types of measurements like DC Current or Resistance measurements. What is the linearity specification for other measurements?

Solution

The linearity is only listed under DC Voltage, but it is listed as ADC Linearity. Since linearity is an ADC specification, it will hold for other types of measurement as well.