Error -225130 in NI-DCPower

Updated Oct 7, 2021

Reported In

Hardware

  • Power Supply
  • PXI Programmable Power Supply

Software

  • LabVIEW

Driver

  • NI-DCPower

Issue Details

I get error -225130 at niDCPower Initiate.vi with the following information when I:

  • Run my NI-DCPower example in LabVIEW
  • Trying to control my SMU in Instrument Studio
  • Try to Self- Calibrate my SMU in NI MAX

Error -225130 occurred at niDCPower Initiate.vi

Possible reason(s): 
Internal Software Error occurred in Power_Supply software. Please contact National Instruments Support. 

Device: <device_name>

Component Name: nipsdk 
File Name: PowerSupply\psd\trunk\17.1\source\nipsd\primitives\dma\input\tChinchInputTransferEnginePrimitive.cpp
Line Number: 292 

Source: 
unknown (0): ??? 
unknown (0): ??? 
unknown (0): ??? 
unknown (0): ??? 
unknown (0): ??? 
unknown (0): ??? 
unknown (0): ??? 
unknown (0): ??? 
unknown (0): ??? 
unknown (0): ??? 


Status Code: -225130 

Solution

Take the following steps to resolve this issue:
  1. Make sure NI-DCPower driver is installed. 
  2. Restart the controller or chassis.
  3. Move SMU/Power Supply to a different slot in the chassis.
  4. If connected through MXI, move the PCI/PCIe card to a 75W slot on the PC's motherboard.
  5. Reinstall NI-DCPower driver to fix a possible corrupt installation. 
  6. Update thunderbolt drivers if using a thunderbolt controller.
  7. Update PXI chassis firmware.
  8. Update controller BIOS.
  9. Update PC motherboard BIOS.
  10.  If the above steps are all performed and yet does not work, your mainboard bios is probably incompatible with the SMU you are using. To restore functionality please install the MXI-Express Bios compatibility tool by referring to the steps in the Readme provided.
If none of the above methods work, please contact your local National Instruments support.