Error -300211 during Verify AC Amplitude Accuracy of PXIe-5162 (4CH) Procedure in Calibration Executive 4.0

Updated May 3, 2018

Reported In

Hardware

  • PXIe-5162

Software

  • Calibration Executive 4.0

Issue Details

I get the following error message during Verify AC Amplitude Accuracy in the PXIe-5162 (4CH) procedure in Calibration Executive 4.0:

Error or termination occurred. This calibration may not be valid.
Error code: -300211
Error message: Error -300211 occurred at cex_ConvertUnitsPrefixes.vi

Solution

To replace the database table, download NI 516x.zip from the Attachments section below, and extract the NI 516x.mdb file to the following location, replacing the old file:

32-bit Operating System: 
C:\Program Files\National Instruments\Calibration Executive\procedures\NI 516x\Limits\... 

64-bit Operating System: 
C:\Program Files (x86)\National Instruments\Calibration Executive\procedures\NI 516x\Limits\...

Additional Information

The database table for the PXIe-5162 Verify AC Amplitude Accuracy test is corrupt and must be replaced. 