Archived:Verify Timebase Accuracy Database Error in Calibration Executive 3.6.4

Updated Mar 20, 2018

National Instruments does not actively maintain this document.

This content provides support for older products and technology, so you may notice outdated links or obsolete information about operating systems or other relevant products.

Reported In

Hardware

  • PXIe-5160

Software

  • Calibration Executive 3.6.4

Issue Details

Why does the PXIe-5160 procedure in Calibration Executive 3.6.4 throw a database error during the Verify Timebase Accuracy test?

Solution

The database table for the PXIe-5160 Verify Timebase Accuracy test is corrupt and must be replaced. 

To replace the database table, download NI 516x.zip from the Attachments section below, and extract the NI 516x.mdb file to the following location, replacing the old file:

32-bit Operating System: 
C:\Program Files\National Instruments\Calibration Executive\procedures\NI 516x\Limits\... 

64-bit Operating System: 
C:\Program Files (x86)\National Instruments\Calibration Executive\procedures\NI 516x\Limits\...

Attachments