Issue Details

I receive Error 200463 when using the NI ELVISmx Digital Writer. Why might I be seeing this?

Solution

Ensure that the number of Digital Output Lines matches the number of Lines to Write. For example, if you choose lines 0-7 for Lines to Write, you could write a 1D array of booleans of length 8 to DO Lines.

See the reference page of the NI ELVISmx Digital Writer for parameter details.

 