How to Create My MP500 Script With Type-F Polling Command in Every 10ms

Updated Mar 26, 2020

Reported In

Hardware

  • Micropross Accessories

Other

H/W : MP500TCL3 OR MP500PT1
S/W : MPManager

Issue Details

  1. I would like to make the script to send Type-F Polling command in every 10ms. 
  2. How to set 1st polling command after 10ms from carrier is enabled

Solution

In order to get exact delay value (e.g. FDT,delay),  we should use script command inside Card/Reader Sequence.
  1. Make sure use Open/Close Sequence
  2. Put the delay 10ms between each command 
  3. Put the delay 10ms after Power Off/ On sequencer