Deployment Error -200431 When Adding CI-Pulse Measurement Channel in VeriStand

Updated Aug 28, 2019

Reported In

Hardware

  • PXI-8110
  • PXI-6251

Software

  • VeriStand
  • LabVIEW

Driver

  • NI-DAQmx

Issue Details

I created a CI-Pulse Measurement channel in VeriStand's system Explorer windows in order to measure Pulse Width Modulation (PWM) in VeriStand. As hardware for the pulse measurement, I used the PXI-6251 model. However, the following error occurred when deploying VeriStand.

The VeriStand Gateway encountered an error while deploying the System Definition file.
Details:
Error -200431 occurred at Project Window.lvlib:Project Window.vi >> Project Window.lvlib:Command Loop.vi >> NI_VS Workspace ExecutionAPI.lvlib:NI VeriStand - Connect to System.vi
Possible reason(s):
Selected physical channel does not support the measurement type required by the virtual channel you are creating.
Create a channel of a measurement type that is supported by the physical channel, or select a physical channel that supports the measurement type.

 

Solution

VeriStand offers the CI-Pulse Measurement type, which is only supported on X Series DAQ devices. The PXI-6251 you tried to use in VeriStand is the M series DAQ device. Therefore, NI recommends you select another X Series DAQ device to use the CI-Pulse Measurement in VeriStand.