Solution
The NI 9237 has two internal resistors for Half-Bridge completion. To connect a strain gage (R4) in the Quarter-Bridge configuration, you need an external precision resistor (R3) to complete the bridge circuit. Wire the strain gage and the precision resistor as shown. For shunt calibration, connect wires across the precision resistor as shown.
The list of pin connections between the NI 9949 and the NI 9237 can be found below:
| NI 9949 | NI 9237 Terminals |
| Pin 1 | SC | Shunt Resistor end 1 |
| Pin 2 | AI+ | Channel Input + |
| Pin 3 | AI- | Channel Input - |
| Pin 4 | RS+ | Remote Sense + |
| Pin 5 | RS- | Remote Sense - |
| Pin 6 | EX+ | Excitation + |
| Pin 7 | EX- | Excitation - |
| Pin 8 | T+ | TEDS+ |
| Pin 9 | T- | TEDS- |
| Pin 10 | SC | Shunt Resistor end 2 |