Is There a Manual Calibration for USB 6501?

Updated Dec 15, 2017

Reported In

Hardware

  • USB-6501

Issue Details

Can the USB 6501 be calibrated manually?

Solution

There are no calibration services offered for the USB 6501. Calibration is limited to verification of the onboard oscillator's frequency. To perform this verification, you will need to measure the onboard clock to see if it is within the specifications of operating frequency, steps can be found below.
  1. Set up a digital output task in LabVIEW and route a counter to the PFI line output that you are reading. 
  2. View the counter output frequency read. 
  3. Make sure that the frequency reading is within the specified operating range. 