6229 Fuse Blown

Updated Nov 21, 2017

Reported In

Hardware

  • PCI-6229
  • PXI-6229

Issue Details

  • The fuse is blown on my 6229
  • What can I do to fix this?

Solution

  • The 6229 manual states the following
    • Older revisions have a self-resetting fuse that opens when current exceeds this specification
    • Newer revisions have a traditional fuse that opens when current exceeds this specification
    • This fuse is not customer-replaceable; if the fuse permanently opens, return the device to NI for repair

Additional Information

WAS THIS ARTICLE HELPFUL?

Not Helpful